Function Gen says 1V, Scope says 2V. Which one's right?
Click to subscribe! ► http://bit.ly/Scopes_Sub ◄
The function generator is set to have a 1 V output, but the oscilloscope measures a 2 Volt signal. Why? When set to 50 ohm mode, the function generator actually outputs double the specified input voltage. It does this because it expects to see a 50 ohm load on the output, creating a 50/50 voltage divider circuit.
Questions or comments? Ask me on Twitter:
@DanielBogdanoff / danielbogdanoff
Learn more about using oscilloscopes:
http://oscilloscopelearningcenter.com
Check out the EEs Talk Tech podcast:
• EEs Talk Tech - An Electrical Enginee...
The digital storage oscilloscope we used: http://www.keysight.com/en/pcx-x20520...
#oscilloscope #oscilloscopes #electronics #electricalengineering #impedance #impedancematching #50ohm #50ohmimpedance #highimpedance #outputimpedance #functiongenerator
Смотрите видео Don't Fall Into This Output Impedance Trap! онлайн без регистрации, длительностью часов минут секунд в хорошем качестве. Это видео добавил пользователь Keysight Labs 11 Сентябрь 2018, не забудьте поделиться им ссылкой с друзьями и знакомыми, на нашем сайте его посмотрели 38,546 раз и оно понравилось 1 тысяч людям.